ESTATE OF MICHAEL CHESTER et al v. BOSTON SCIENTIFIC CORP. et al
Filing
18
OPINION filed. Signed by Judge Freda L. Wolfson on 2/27/2017. (km)
**NOT FOR PUBLICATION**
UNITED STATES DISTRICT COURT
DISTRICT OF NEW JERSEY
______________________________
BARBARA A. CHESTER, as
Administratrix ad Prosequendum of
Estate of Michael Chester, and
BARBARA A. CHESTER, as an
Individual,
Plaintiff,
v.
BOSTON SCIENTIFIC CORP.,
GUIDANT CORPORATION,
GUIDANT LLC, CARDIAC
DEFIBRILLATORS, INC.,
Defendants.
:
:
:
:
:
:
:
:
:
:
:
:
:
:
:
:
:
:
Civil Action No. 16-02421 (FLW)
OPINION
WOLFSON, United States District Judge:
Before the Court is the motion of Defendants Boston Scientific, Corp., Guidant
Corporation, Guidant LLC, and Cardiac Defibrillators, Inc., to dismiss the product liability and
common law tort claims of Plaintiff Barbara A. Chester. Plaintiff contends that her deceased
husband, Michael Chester, suffered harm and died due to the defective connective wires in the
defibrillator device designed, manufactured, and sold by Defendants. For the reasons set forth
below, the Court finds that the New Jersey Products Liability Act (“PLA”) subsumes Plaintiff’s
state law tort claims into her strict product liability claim, and that Plaintiff’s strict product
liability claim is, in turn, preempted by the Medical Device Amendments (“MDA”) to the
Federal Food, Drug, and Cosmetic Act (“FDCA”). Accordingly, Defendants’ motion to dismiss
is granted.
1
FACTUAL BACKGROUND & PROCEDURAL HISTORY
Plaintiff was the lawful wife of Michael Chester. Compl. ¶ 11. In or about May 2004,
Michael Chester, who suffered from cardiac problems, underwent a surgical procedure to have a
defibrillator device implanted in his chest to treat his heart condition. Id. at ¶ 12. In or about May
2012, Mr. Chester, who continued to suffer from cardiac problems, underwent an additional
surgical procedure to implant another defibrillator device. Id. at ¶ 13. The devices that were
finally implanted after both surgeries were (1) a Boston Scientific Incepta CRT-D Model # N164
Serial # 100005 defibrillator device and (2) an Endotak Edurance EZ lead, Model # 0154 Serial #
354220 (collectively, the “Defibrillator” or the “device”). Id. at ¶ 15. Defendants Boston
Scientific Corp., Guidant Corporation, and Guidant LLC, and Cardiac Defibrillators, Inc.,
designed, manufactured, fabricated, supplied, and sold the Defibrillator. Id. at ¶ 16.
In October 2013, Mr. Chester underwent another defibrillator implantation surgery at the
University of Pennsylvania. Id. at ¶ 20. An unidentified representative of Defendant Boston
Scientific was present in the operating room during the implantation surgery. Id. at ¶ 21. After
the implantation surgery, a representative of Defendant Boston Scientific informed Plaintiff and
Mr. Chester’s daughter, acting as his next of kin, that Boston Scientific was having a “newer”
and “better” model of defibrillator implanted in Mr. Chester, because the prior Defibrillator was
defective. Id. at ¶ 20. The representative, stated that the “leads” or wires connecting the device to
Mr. Chester’s heart, were known to “slip.” Id. at ¶¶ 28-29. The representative informed Plaintiff
and Mr. Chester’s daughter that this propensity of the leads to slip was the reason why the
October 2013 surgery was necessary to remove the leads from Mr. Chester. Id. at ¶ 29. 1
1
The Court notes that a slightly different version of the conversation with the Boston Scientific
representative appears in the supplemental certification of Barbara Ann Eisenstein, Mr. Chester’s
daughter, which was submitted in the state court proceeding before removal in support of the
2
On June 2, 2014, Mr. Chester died. Id. at ¶ 23. Plaintiff alleges that, due to the defective
leads, the device was unfit for the purpose for which it was sold — the treatment of heart failure
— and that the device, in an unspecified manner, caused Mr. Chester serious physical harm and
lead to his death. 2 Id. at ¶¶ 47-49.
On October 8, 2015, Plaintiff filed the Original Complaint in this case in the Superior
Court of New Jersey, Law Division, Ocean County. The Original Complaint contained eight (8)
Counts and named Defendants, as well as hospitals, healthcare providers, and fictitious entities
(collectively the “Medical Provider Defendants”). Defendants moved to dismiss on MDA
preemption and other grounds, and, on January 27, 2016, Plaintiff cross-moved for leave to
amend the Original Complaint to cure her pleading defects. In mid-April 2016, Plaintiff’s claims
against the Medical Provider Defendants were dismissed by a series of stipulations of the parties,
resulting in diversity of citizenship among the parties. Defendants removed the case to this Court
on April 29, 2016, before the state court adjudicated the motion to dismiss and cross-motion for
leave to amend.
Original Complaint and again in opposition to the motion. The version of events in the Amended
Complaint is reproduced in this Opinion and controls for purposes of this motion. On a motion to
dismiss, only the pleadings, and not the supplementary certifications of the parties, may be
considered. In either case, the fact of the conversation is immaterial to the Court’s decision.
2
The Court notes the unorthodox nature of the Complaint in this products liability action in
failing to specify the manner in which the allegedly defective device actually harmed Mr.
Chester or caused his death. As Defendants observe in their briefing, no causal connection
between the allegedly defective device and Mr. Chester’s alleged injuries and death is alleged.
To the contrary, the complaint alleges that the device was removed from Mr. Chester in October
2013, seven months before his June 2014 death. Far from alleging that faulty connective wires
harmed or killed Mr. Chester, therefore, the Complaint actually suggests that the wires were
removed and replaced far in advance of his death. Although the Defendants move for dismissal
on the basis of this and other pleading deficiencies under Rules 8 and 9, the Court need not reach
those arguments as dismissal is warranted on other bases.
3
On May 26, 2016, this Court entered a Consent Order granting Plaintiff leave to file an
Amended Complaint in substantial conformity with the proposed Amended Complaint submitted
with Plaintiff’s cross-motion for leave to amend that had been filed in the New Jersey Superior
Court. On June 24, 2016, Plaintiff filed her Amended Complaint, the Complaint currently before
the Court on Defendants’ motion. The Amended Complaint presents five (5) claims against
Defendants, claiming wrongful death and survival damages arising from Plaintiff’s husband, Mr.
Chester’s injury and death due to the alleged design, manufacturing, and labeling defect of
Defendants’ Defibrillator. In Count I, Plaintiff raises a strict product liability failure to warn
claim under the New Jersey Products Liability Act (“PLA”), which, through its paragraph
incorporating prior allegations by reference, also appears to raise a breach of implied warranty
claim. Compl. ¶¶ 17, 50-62. In Count II, Plaintiff raises common law consumer fraud, deceit, and
fraudulent concealment claims, again based on Defendants’ alleged failure to warn. Id. at ¶¶6373. In Count III, Plaintiff raises a common law claim for negligent misrepresentation. Id. at ¶¶
74-80. In Count IV, Plaintiff raises a common law claim for intentional infliction of emotional
distress. Id. at ¶¶ 81-85. Finally, in Count V, Plaintiff raises a common law claim for loss of
consortium. Id. at ¶¶ 86-89.
Defendants moved to dismiss the Amended Complaint on July 8, 2016, on the grounds
that (i) Plaintiff’s state common law claims are subsumed by the PLA into her Count I strict
liability failure to warn claim and are not separately actionable; (ii) the Medical Device
Amendments to the Federal Food, Drug, and Cosmetics Act expressly and impliedly preempt
both Plaintiff’s PLA and common law claims; and (iii) Plaintiff fails to meet the pleading
standards of Federal R. Civ. P. 8 and Rule 9(b). Plaintiff opposed the motion on July 27, 2016.
Plaintiff’s Opposition does not respond to Defendants’ PLA subsumption argument, but contends
4
that Plaintiff’s claims are “parallel claims” falling within a recognized exception to MDA
preemption, that Plaintiff has adequately pleaded her claims, and that, in the alternative, Plaintiff
should be granted discovery to obtain additional facts to support her claims.
JURISDICTION
The Court has subject matter jurisdiction over this matter pursuant to 28 U.S.C. § 1332
because the amount in controversy is greater than $75,000, and there is complete diversity of the
parties. The parties do not contest subject matter or personal jurisdiction, and agree that venue is
appropriate in this Court.
STANDARD OF REVIEW
Federal Rule of Civil Procedure 12(b)(6) provides that a court may dismiss a claim “for
failure to state a claim upon which relief can be granted.” 3 When reviewing a motion to dismiss,
courts must first separate the factual and legal elements of the claims, and accept all of the wellpleaded facts as true. See Fowler v. UPMC Shadyside, 578 F.3d 203, 210-11 (3d Cir. 2009). All
reasonable inferences must be made in the plaintiff’s favor. See In re Ins. Brokerage Antitrust
Litig., 618 F.3d 300, 314 (3d Cir. 2010). In order to survive a motion to dismiss, the plaintiff
must provide “enough facts to state a claim to relief that is plausible on its face.” Bell Atl. Corp.
3
The Third Circuit has held that “because federal preemption is an affirmative defense on which
the defendant bears the burden of proof[,] . . . a motion under Rule 12(c) for judgment on the
pleadings is a more appropriate procedural vehicle for dismissing cases on preemption grounds,
instead of a motion under Rule 12(b)(6), except for cases in which preemption is manifest in the
complaint itself.” In re Asbestos Prod. Liab. Litig. (No. VI), 822 F.3d 125, 1330 n.6 (3d Cir.
2016). The Court will apply the standard under Rule 12(b)(6) in this case because all of the facts
assertedly giving rise to federal preemption are present in the Complaint itself. In any event,
application of the other standard in Rule 12(c) would not affect the outcome in this case because
“[i]n deciding a motion under Rule 12(c), the court must view the facts presented in the
pleadings and the inferences to be drawn therefrom in the light most favorable to the nonmoving
party, and may not grant the motion unless the movant clearly establishes that no material issue
of fact remains to be resolved and that he is entitled to judgment as a matter of law.” Id.
(citations omitted).
5
v. Twombly, 550 U.S. 544, 570 (2007). This standard requires the plaintiff to show “more than a
sheer possibility that a defendant has acted unlawfully,” but does not create as high of a standard
as to be a “probability requirement.” Ashcroft v. Iqbal, 556 U.S. 662, 678 (2009).
The Third Circuit requires a three-step analysis to meet the plausibility standard
mandated by Twombly and Iqbal. First, the court should “outline the elements a plaintiff must
plead to a state a claim for relief.” Bistrian v. Levi, 696 F.3d 352, 365 (3d Cir. 2012). Next, the
court should “peel away” legal conclusions that are not entitled to the assumption of truth. Id.;
see also Iqbal, 556 U.S. at 678-79 (“While legal conclusions can provide the framework of a
complaint, they must be supported by factual allegations.”). It is well-established that a proper
complaint “requires more than labels and conclusions, and a formulaic recitation of the elements
of a cause of action will not do.” Twombly, 550 U.S. at 555 (internal quotations and citations
omitted). Finally, the court should assume the veracity of all well-pled factual allegations, and
then “determine whether they plausibly give rise to an entitlement to relief.” Bistrian, 696 F.3d at
365 (quoting Iqbal, 556 U.S. at 679). A claim is facially plausible when there is sufficient factual
content to draw a “reasonable inference that the defendant is liable for the misconduct alleged.”
Iqbal, 556 U.S. at 678. The third step of the analysis is “a context-specific task that requires the
reviewing court to draw on its judicial experience and common sense.” Id. at 679.
ANALYSIS
A. Subsumption of Tort Claims Under the PLA 4
4
Plaintiff does not provide any arguments against PLA subsumption in her Opposition, and
indeed may have waived her common law claims. See Opposition at 13 (“Plaintiff’s claims are
premised, inter alia, on the Product Liability Act . . . which governs all claims for harm caused
by a product other than breach of an express warranty.”). As Plaintiff’s arguments elsewhere in
her Opposition suggest she did not intend to waive her claims, despite failing to oppose
Defendants’ motion on the basis of subsumption, the Court nevertheless analyzes the application
of the PLA to these claims at length.
6
“The PLA ‘established the sole method to prosecute a product liability action’ such that
‘only a single product liability action remains.’” Clements v. Sanofi-Aventis, U.S., Inc., 111 F.
Supp. 3d 586, 596 (D.N.J. 2015) (quoting Tirrell v. Navistar Int'l, Inc., 248 N.J. Super. 390,
398–99, (App. Div. 1991)). See also In re Avandia Mktg. Sales Practices & Prod. Liab. Litig.,
588 F. App'x 171, 178 (3d Cir. 2014) (same). “The language chosen by the Legislature in
enacting the PLA is both expansive and inclusive, encompassing virtually all possible causes of
action relating to harms caused by consumer and other products.” In re Lead Paint Litig., 191
N.J. 405, 436–47 (2007). It “effectively creates an exclusive statutory cause of action for claims
falling within its purview.” Repola v. Morbark Indus., Inc., 934 F.2d 483, 492 (3d Cir. 1991).
See also id. (correctly anticipating that “the New Jersey Supreme Court would hold that the
NJPLA generally subsumes common law product liability claims, thus establishing itself as the
sole basis of relief under New Jersey law available to consumers injured by a defective
product.”); Sinclair v. Merck & Co., 195 N.J. 51, 65 (2008) (finding “the clear intention of our
Legislature to include all [product liability] claims within the scope of the PLA”) (quotation
omitted, alterations in original).
Under the PLA, a product liability action is defined as “any claim or action brought by a
claimant for harm caused by a product, irrespective of the theory underlying the claim, except
actions for harm caused by breach of an express warranty.” N.J.S.A. 2A:58C–1(b)(3) (emphasis
added). 5 In determining whether a tort claim is subsumed under the PLA, the New Jersey
Supreme Court has observed that “[t]he essential question is whether plaintiffs' effort to recover .
5
The PLA contains two exceptions for causes of action for breach of an express warranty and
environmental torts, neither of which is applicable in this case. Sinclair, 195 N.J. at 62.
7
. . damages is limited by the definition of ‘harm’ in the PLA.” Sinclair, 195 N.J. at 62. “Harm” is
defined in the PLA as
(a) physical damage to property, other than to the product itself; (b) personal physical
illness, injury or death; (c) pain and suffering, mental anguish or emotional harm; and (d)
any loss of consortium or services or other loss deriving from any type of harm described
in subparagraphs (a) through (c) of this paragraph.
N.J.S.A. 2A:58C–1(b)(2). “In short, those former common-law causes of action . . . have merged
into a single cause of action under the PLA,” and thus are no longer recognized “as viable
separate claims for harm deriving from a defective product.” Clements, 111 F. Supp. 3d at 596.
Courts in this Circuit therefore consistently dismiss product liability claims based on those
common-law theories. Id. at 597 n. 5 (collecting cases).
Here, all five Counts of the Complaint, although presenting different common law
theories, allege the same harm caused by Defendants’ defective product. Count I alleges:
As a direct and proximate result of the foregoing, . . . [Plaintiffs] were caused to sustain
the loss of income, support, society, love, grief, consortium, services, guidance, care,
comfort, companionship, and inheritance of the decedent and suffered mental anguish,
mental pain and suffering, emotional distress, and were caused to incur other necessary
and reasonable expenses as a result of the decedent’s death, including but not limited to,
funeral and burial costs, medical expenses, and were otherwise damaged.
Compl. ¶ 61. Counts II, III, and IV incorporate the exact same language by reference. Compl. ¶¶
72, 79, 84. Count V explicitly seeks damages only for loss of consortium. Compl. ¶ 89. As these
harms fall within the PLA, the common law causes of action in Count I and the entirety of
Counts II, III, IV, and V, which allege only common law causes of action, are therefore
subsumed by the PLA.
Accordingly, Count I is dismissed to the extent that it presents an implied warranty
(Compl. ¶ 17, incorporated into Count I by ¶ 50) theory of liability. See, e.g., Mendez v. Shah, 28
F. Supp. 3d 282, 294 (D.N.J. 2014) (implied warranty claim subsumed by PLA). Count II
8
(consumer fraud, deceit, and fraudulent concealment) is dismissed. See Delaney v. Stryker
Orthopaedics, No. CIV.A. 08-03210DMC, 2009 WL 564243, at *7 (D.N.J. Mar. 5, 2009)
(consumer fraud claim subsumed by PLA); Sinclair, 195 N.J. at 66 (same). Count III (negligent
misrepresentation) is dismissed. See Calender v. NVR Inc., 548 F. App'x 761, 764 (3d Cir. 2013)
(“The PLA does not recognize negligence or implied breach of warranty as separate claims for
harm caused by a product. Rather, the PLA is the exclusive remedy for such actions and other
claims are subsumed within the statutory cause of action.”). Count IV (intentional infliction of
emotional distress) is dismissed. See N.J.S.A. 2A:58C–1(b)(2)(c) (subsuming product actions for
“pain and suffering, mental anguish or emotional harm”). And Count V (loss of consortium) is
dismissed. See id. at 2A:58C–1(b)(2)(d) (subsuming product actions for “loss of consortium or
services or other loss deriving from any type of harm described in subparagraphs (a) through
(c)”).
The only claim in Plaintiff’s complaint properly brought, pursuant to the PLA, is
Plaintiff’s Count I strict product liability claim for failure to provide adequate warning.
Defendant contends that, even if proper under state law, Plaintiff’s failure to warn claim
nevertheless must be dismissed as preempted by the federal MDA. This Court agrees.
B. Preemption Under the MDA
In Count I, Plaintiff alleges that Defendants failed to adequately warn of the substantial
dangers of the Defibrillator, due to defects in the design, manufacture, and distribution of the
device and its components, namely the Endotak Edurance EZ Leads, Model # 0154 Serial #
354220. Compl. ¶ 55. Defendants’ failure to warn is alleged to have two constituent parts. First,
Defendants are alleged to have “failed to provide adequate warnings, instructions, guidelines, or
admonitions to members of the public, including decedent [Mr. Chester], of the design and
9
manufacturing defects, which defendants knew, or in the exercise of reasonable care should have
known, to have existed in the defibrillator device and leads.” Id.at ¶ 57. Second, Defendants are
alleged to have “had a continuing obligation to notify the FDA of product defects causing serious
injury or death, and defendants failed to abide by this federally mandated obligation.” Id. at ¶ 56.
As explained below, both parts are essential to Plaintiff’s theory of liability, but, even together,
do not save Plaintiff’s claim from federal preemption.
“The Federal Food, Drug, and Cosmetic Act (FDCA), 52 Stat. 1040, as amended, 21
U.S.C. § 301 et seq., has long required FDA approval for the introduction of new drugs into the
market.” Riegel v. Medtronic, Inc., 552 U.S. 312, 315 (2008). It did not, however, originally
govern medical devices. “[T]he introduction of new medical devices was left largely for the
States to supervise as they saw fit.” Ibid.
Congress “swept back some state obligations and imposed a regime of detailed federal
oversight” with the passage of the Medical Device Amendments of 1976, 21 U.S.C. § 360c et
seq. (MDA). Id. at 316. “The devices receiving the most federal oversight are those in Class III,
which include replacement heart valves, implanted cerebella stimulators, and pacemaker pulse
generators.” Riegel, 552 U.S. at 317. Under the MDA, such devices are subject to a regime of
“premarket approval” (“PMA”) by the FDA before they can be marketed and sold to the public.
“Premarket approval is a ‘rigorous’ process.” Id. at 317 (quoting Medtronic, Inc. v. Lohr, 518
U.S. 470, 477 (1996)). The FDA “grants premarket approval only if it finds there is a ‘reasonable
assurance’ of the device’s ‘safety and effectiveness.’” Id. at 317 (quoting 21 U.S.C. § 360e(d)).
“The agency must ‘weig[h] any probable benefit to health from the use of the device against any
probable risk of injury or illness from such use.’ It may thus approve devices that present great
10
risks if they nonetheless offer great benefits in light of available alternatives.” Id. at 318 (quoting
§ 360c(a)(2)(C)).
“The premarket approval process includes review of the device’s proposed labeling. The
FDA evaluates safety and effectiveness under the conditions of use set forth on the label, §
360c(a)(2)(B), and must determine that the proposed labeling is neither false nor misleading, §
360e(d)(1)(A).” Id. at 318. “Once a device has received premarket approval, the MDA forbids
the manufacturer to make, without FDA permission, changes in design specifications,
manufacturing processes, labeling, or any other attribute, that would affect safety or
effectiveness. § 360e(d)(6)(A)(i). If the applicant wishes to make such a change, it must submit,
and the FDA must approve, an application for supplemental premarket approval, to be evaluated
under largely the same criteria as an initial application. § 360e(d)(6); 21 CFR § 814.39(c).” Id. at
319.
“After premarket approval, the devices are subject to reporting requirements. § 360i.
These include the obligation to inform the FDA of new clinical investigations or scientific
studies concerning the device which the applicant knows of or reasonably should know of, 21
CFR § 814.84(b)(2), and to report incidents in which the device may have caused or contributed
to death or serious injury, or malfunctioned in a manner that would likely cause or contribute to
death or serious injury if it recurred, § 803.50(a).” Id. at 319.
1. Express Preemption
The MDA includes an express pre-emption provision that states:
Except as provided in subsection (b) of this section, 6 no State or political subdivision of a
State may establish or continue in effect with respect to a device intended for human use
any requirement—
6
The exception contained in subsection (b) permits the FDA to exempt some state and local
requirements from pre-emption and is not at issue here.
11
(1) which is different from, or in addition to, any requirement applicable under this
chapter to the device, and
(2) which relates to the safety or effectiveness of the device or to any other matter
included in a requirement applicable to the device under this chapter.”
21 U.S.C. § 360k(a). “In Riegel v. Medtronic, Inc., the Supreme Court held that state-imposed
requirements are preempted by the MDA if (1) the Federal Government has established
requirements applicable to the device and (2) the plaintiff's claims are based on state
requirements related to safety and effectiveness that are ‘different from, or in addition to’ the
federal requirements.” Williams v. Cyberonics, Inc., 388 F. App'x 169, 171 (3d Cir. 2010)
(quoting Riegel, 552 U.S. at 321-22). “Riegel concluded that the first prong of the preemption
test is automatically satisfied where a medical device has received premarket approval. The PMA
process necessarily imposes federal requirements, the Court found, because the FDA requires
that a medical device be ‘made with almost no deviations from the specifications in [the PMA]
application.’” Clements, 111 F. Supp. 3d at 597 (quoting Riegel, 552 U.S. at 321-23. See also
Cornett v. Johnson & Johnson, 211 N.J. 362, 387 (2012) abrogated on other grounds by
McCarrell v. Hoffmann–La Roche, Inc., No. 076524, 2017 WL 344449 (N.J. Jan. 24, 2017)
(where the product “underwent the rigorous and individualized PMA process for Class III
medical devices . . . [t]he approval provided by the FDA communicates that defendants
demonstrated the safety and effectiveness of the product for its approved uses. Moreover, that
approval includes the label and instructions that accompany the device. The totality of the
approval represents a specific federal requirement.”).
The Defibrillator and its components are Class-III medical devices that were required to
undergo and have received Premarket Approval from the FDA conditioned upon continued
12
compliance with FDA regulations. Compl. ¶ 31. 7 Under Riegel, prong one of the preemption
analysis is met. The only question before the Court, therefore, is whether the state requirements
imposed by Plaintiff’s failure to warn PLA claim relate to the “safety and effectiveness of the
device” and are “different from, or in addition to” the federal requirements of the PMA process.
In passing the PLA, the New Jersey “Legislature intended . . . to limit the liability of
manufacturers so as to balance the interests of the public and the individual with a view towards
economic reality.” Sinclair, 195 N.J. at 62. Plaintiff’s product liability claim under New Jersey
law is thus subject to a strict limitation on liability.
A manufacturer or seller of a product shall be liable in a product liability action only if
the claimant proves by a preponderance of the evidence that the product causing the harm
was not reasonably fit, suitable or safe for its intended purpose because it: a. deviated
from the design specifications, formulae, or performance standards of the manufacturer
or from otherwise identical units manufactured to the same manufacturing specifications
or formulae, or b. failed to contain adequate warnings or instructions, or c. was designed
in a defective manner.
N.J.S.A. 2A:58C-2.
As relevant to this action, the PLA further provides:
In any product liability action the manufacturer or seller shall not be liable for harm
caused by a failure to warn if the product contains an adequate warning or instruction or,
in the case of dangers a manufacturer or seller discovers or reasonably should discover
after the product leaves its control, if the manufacturer or seller provides an adequate
warning or instruction.
N.J.S.A. 2A:58C–4. Accordingly, “[i]n a failure-to-warn case, ‘the duty to warn is premised on
the notion that a product is defective absent an adequate warning for foreseeable users that the
7
Although the Court relies only upon the allegations in the Complaint, it also notes that the
parties do not dispute that the Defibrillator and its components are Class-III medical devices that
have received Premarket Approval from the FDA. See, e.g., Opposition at 5 (Defendants’
contention on the motion that “the claims against movants concern a Class III medical device,
regulated by the Food and Drug Administration (‘FDA’), and approved through the FDA’s PreMarket Approval (‘PMA’) process.”); id. at 16 (Plaintiff requesting discovery into the PMA
process on the Defibrillator, attaching documents showing PMA status of Defibrillator).
13
product can potentially cause injury.’” Mendez v. Shah, 28 F. Supp. 3d 282, 299 (D.N.J. 2014)
(quoting Clark v. Safety–Kleen Corp., 179 N.J. 318, 336 (2004)). “The plaintiff must establish
that the defendant had a duty to warn, and then establish that an adequate warning was not
provided.” Id. (citing James v. Bessemer Processing Co., 155 N.J. 279 (1998)). “Plaintiff must
then prove the breach of duty (the absence of a warning) was a proximate cause of the accident.”
(citing Coffman v. Keene Corp., 133 N.J. 581 (1993)).
Under the PLA then, a manufacturer is liable for failure to warn only if the plaintiff can
prove that the device was defective — meaning not fit, suitable, or safe for its intended purpose
absent the desired warning, N.J.S.A. 2A:58C-2 — and that the warning was not actually
provided, N.J.S.A. 2A:58C–4. These state requirements impose different and additional
obligations on Defendants than the federal obligations of the PMA process. In granting the
Defibrillator PMA, the FDA already evaluated the safety and effectiveness of the device under
the conditions set forth on the label and determined that the labeling was not false or misleading.
Riegel, 552 U.S. at 318 (citing 21 U.S.C. §§ 360c(a)(2)(B), 360e(d)(1)(A)). To find that the
device is unsafe and defective because of the omission of a warning from the FDA-approved
label would therefore overturn the FDA’s determination and impose a stricter state standard. The
MDA does not permit such additional and more stringent state requirements, and so Plaintiff’s
PLA claim is dismissed. Horn v. Thoratec Corp., 376 F.3d 163, 173-177 (3d Cir. 2004) (“The
only state requirements asserted by [plaintiff] are general requirements stemming from state
common law: the HeartMate [device] was designed in a defective manner, it was manufactured
in a defective manner, and the manufacturer had failed to warn of the alleged defects. . . .
Because these state common law claims and duties are in severe tension with § 360k(a) in that
they are either in addition to, or different from, the federal requirements established by the FDA
14
in approving the HeartMate, they are necessarily preempted by federally imposed PMA
requirements under § 360k(a).”). 8
In concluding that Plaintiff’s PLA claim is expressly preempted by the MDA, this Court
is in accord with the only other court in this district to specifically address an alleged PLA claim
arising from defective leads in a defibrillator device. In a closely factually analogous case,
another judge in this District found that
Plaintiffs’ claims of, inter alia, negligence, defective design, and failure to warn stem
from state common law. Plaintiffs would only be able to prevail on the New Jersey PLA
claims if they proved that the lead wire, as designed, manufactured, and distributed, was
defective and unreasonably dangerous. It follows that liability would necessitate a finding
that the lead wire—designed, manufactured, and labeled in a way that the FDA deemed
safe and effective—was both defective and unreasonably dangerous. Such a
determination would necessarily constitute a requirement different from, or in addition to,
the standard required by federal authorities. . . . This is the exact situation when the MDA
requires preemption.
Desai v. Sorin CRM USA, Inc., No. CIV. 12-2995, 2013 WL 163298, at *5 (D.N.J. Jan. 15,
2013) (quotation omitted). This Court’s decision and the decision in Desai, are also consistent
with the overwhelming weight of precedent in this District. See, e.g., Millman v. Medtronic, No.
14-CV-1465, 2015 WL 778779, at *5 (D.N.J. Feb. 24, 2015) (“Because the FDA determined, by
granting PMA, that the Activa System [device] was safe and effective as manufactured and
designed in the PMA application, this Court finds that Plaintiffs' claims based on a
manufacturing defect, design defect, failure to warn, negligence and a breach of
contract/warranty theory are preempted because they impose requirements that are different from
the federal requirements set forth in the PMA process. Each claim, as currently alleged, seeks to
impose requirements that are either different from or in addition to those required by federal
8
The state law claims at issue in Horn were brought under the common law of Pennsylvania, but
were of a type — design defect, manufacturing defect, and failure to warn — that would clearly
have been subsumed by the PLA had they been brought under New Jersey law.
15
law.”); id. at *6 n. 3 (“Plaintiffs do not allege that Medtronic failed to provide any warnings
required by FDA through the PMA process. To the extent Plaintiffs assert that, to comply with
state law, Medtronic was required to give additional or different warnings than those required by
FDA, this claim is preempted.”); Mendez v. Shah, 28 F. Supp. 3d 282, 292 (D.N.J. 2014) (“To
the extent that any of plaintiff's state law claims assert that the warnings on the Infuse/LT–Cage
device were insufficient because they did not warn of dangers . . . such claims are expressly
preempted. To require defendant to add language to the warning would impose an additional
requirement relating the safety or effectiveness of the device.”); Delaney v. Stryker
Orthopaedics, No. CIV.A. 08-03210DMC, 2009 WL 564243, at *3 (D.N.J. Mar. 5, 2009)
(quotations omitted) (“Counts One and Three of Plaintiff's Complaint assert claims for failure to
warn and defective design. . . . As in Riegel, these claims are expressly preempted because they
assert general tort duties of care, allege that a device was designed, labeled, or manufactured in
an unsafe or ineffective manner, and impose different or additional requirements related to the
safety and effectiveness of the Trident™ device.”). 9
9
The Court also notes that the Supreme Court’s holding in Riegel constitutes an alternative and
sufficient basis for the dismissal of Plaintiff’s common law tort claims subsumed by the PLA.
The Riegel Court considered state common law tort claims and found that they too impose state
requirements different from, or in addition to the federal PMA requirements. Id. at 323. It
reasoned that safety and effectiveness are “the very subjects” of such tort causes of action, while
also the primary concerns reserved to the FDA in the federal PMA process. Ibid. As another
court in this District aptly observed:
The upshot of Riegel, then, is that the MDA preempts state tort claims to the extent that
they would impose requirements on device manufacturers that deviate from those
imposed by federal law—particularly, those imposed in the PMA process. Accordingly,
courts within the Third Circuit have consistently held that tort claims based on
negligence, manufacturing and design defects, strict liability, breach of warranty, and
failure to warn, including such claims as subsumed by the New Jersey PLA, are
preempted by MDA.
16
2. No Parallel Claim
In the Amended Complaint and in her Opposition to the motion to dismiss, Plaintiff
contends that, despite directly bearing on issues addressed by the FDA in the PMA process,
Plaintiff’s PLA claims may nevertheless proceed because they fall within a category of state
claims identified in Riegel, which impose requirements “parallel” and not in addition to the
federal requirements of the MDA, and therefore escape express preemption. Plaintiff is correct
that not all state law claims are expressly preempted by the MDA. The Supreme Court has held
that “[s]tate requirements are pre-empted under the MDA only to the extent that they are
‘different from, or in addition to’ the requirements imposed by federal law. § 360k(a)(1). Thus, §
360k does not prevent a State from providing a damages remedy for claims premised on a
violation of FDA regulations; the state duties in such a case ‘parallel,’ rather than add to, federal
requirements.” Riegel, 552 U.S. at 330 (quoting Lohr, 518 U.S., at 495). Accordingly, to fall
within the parallel claim exemption, Plaintiff must allege (i) violation of FDA regulations; (ii) an
independent state law cause of action providing a damages remedy for the violation of FDA
regulations; and (iii) a “cognizable link” between the violation of FDA regulations and Plaintiff’s
injury. Desai, 2013 WL 163298, at *6.
In attempting to establish the first prong, Plaintiff, in the Amended Complaint, provides a
laundry list of FDA regulations with which Defendants were obligated to comply in designing,
manufacturing, and selling the Defibrillator. See Compl. ¶ 36 (requirement to report if a device
Clements, 111 F. Supp. 3d at 598. See also Williams v. Cyberonics, Inc., 388 F. App'x 169, 171
(3d Cir. 2010) (“Generalized common law theories of liability . . . are precisely the type of
claims the MDA sought to preempt.”); ibid. (“Success on appellants' . . . claims would require
them to show that the . . . device was unsafe or ineffective despite the PMA process, thereby
interfering with the requirements already established by the MDA, which has preempted safety
and effectiveness determinations for a device.”).
17
may have caused death or serious injury or device malfunction, and duty to investigate each
adverse event, 21 C.F.R. § 803.50); ¶ 37 (duty to provide detailed descriptions of adverse events
to the FDA, 21 C.F.R. § 803.52); ¶ 38 (duty to conduct trend analysis or undertake remedial
action to prevent unreasonable risk to the public, 21 C.F.R. § 803.53); ¶ 39 (duty to report device
corrections or removals, 21 C.F.R. § 806.10); ¶ 40 (duty to implement design-control systems
and conduct design validation, 21 C.F.R. § 820); ¶ 42 (duty to submit a supplementary report to
the FDA when unanticipated adverse effects of device or device failures require labeling,
manufacturing or design modification, 21 C.F.R. § 814.39). In support of the dismissed common
law counts (all Counts except Count I), Plaintiff also alleges that Defendants are liable pursuant
to 21 C.F.R. § 814.84(b)(2) (requirement that periodic reports to the FDA contain a summary of
information not previously submitted); and 21 C.F.R. § 803.50(a) (obligation of device
manufactures to report to the FDA).
What is missing from Plaintiff’s first prong allegations is any plausible pleading of if,
how, or when Defendants violated any of the listed regulations. Instead, the Amended Complaint
alleges only that “Defendants have a long history of failure to report adverse effects related to
their defibrillator and/or lead devices pursuant to federal law and such was the case here,” id. at ¶
44, and that “Defendants’ failure to meet federal regulations applicable to the medical device at
issue herein, and defendants’ other acts and omissions as described herein directly and
proximately caused the device at issue to be in violation of federal law and unfit for sale,” id. at ¶
47. These conclusory, legal allegations are insufficient to allege a violation of an FDA regulation
sufficient to state a parallel claim. Iqbal, 556 U.S. at 678-79 (“While legal conclusions can
provide the framework of a complaint, they must be supported by factual allegations.”). See also
Clements, 111 F. Supp. 3d at 598 (“The parallel claim exception to preemption, however,
18
requires more than just a change of terminology; a plaintiff cannot simply incant the magic
words [Defendant] violated FDA regulations in order to avoid preemption.” (quotation omitted));
Desai, 2013 WL 163298, at *6–7 (“broad references to federal regulations in pleadings are
insufficient . . . . Allowing a plaintiff to plead non-specific regulations as a basis for a parallel
claim is inconsistent with the Supreme Court's reasoning in Riegel, as well as the pleading
requirements articulated in Twombly, Iqbal, and Fowler.” (quotation omitted)).
Even had Plaintiff pleaded identifiable violations of FDA regulations, Plaintiff’s claim
would still fail at the second prong for failure to identify a state cause of action based on those
violations. The facts of this case are substantially analogous to those in the persuasive decision of
the Eighth Circuit in In re Medtronic, Inc., Sprint Fidelis Leads Prod. Liab. Litig., also dealing
with implanted leads. The complaint there alleged that Defendants
failed to adequately warn consumers of “known defects” and that the . . . [l]eads
presented an unreasonably dangerous risk beyond what the ordinary consumer would
reasonably expect. These claims are preempted by § 360k. The FDA's PMA approval
includes specific language for Class III device labels and warnings. Plaintiffs did not
allege that Medtronic modified or failed to include FDA-approved warnings. Rather, they
alleged that, by reason of state law, Medtronic was required to give additional warnings,
precisely the type of state requirement that is “different from or in addition to” the federal
requirement and therefore preempted.
623 F.3d 1200, 1205 (8th Cir. 2010) (quoting Riegel, 552 U.S. at 330). In short, Plaintiff here
was required to allege that the violation of the FDA reporting requirements listed in the
Amended Complaint was the basis of her state law cause of action. This Plaintiff failed to do,
instead alleging that the Defibrillator’s FDA-approved warnings, rather than violating FDA
reporting requirements, violated the New Jersey PLA by failing to provide adequate notice of
danger. See Compl. ¶ 59 (“Because information concerning the inherently dangerous
characteristics of the subject devices were unknown to and concealed from the general public
and to the decedent, . . . defendants were under a duty, yet intentionally failed to disclose to
19
patients, the consuming public, and the medical community, that their product was defective,
unsafe, and inherently dangerous for their intended use by consumers such as the decedent.”).
Such claims are expressly preempted as explained above.
Although Defendants also argue that to the extent Plaintiff alleges a violation of FDA
regulations, her claims based upon such violations are impliedly preempted as impermissible
attempts to enforce FDA reporting requirements under the Supreme Court’s decision in Buckman
Co. v. Plaintiffs' Legal Comm., 531 U.S. 341 (2001), the Court does not read the Amended
Complaint as asserting any “fraud-on-the-FDA” claim based on the alleged reporting regulations.
The MDA does, however, provide that all actions to enforce FDA requirements “shall be by and
in the name of the United States,” 21 U.S.C. § 337(a), and, in Buckman, the Supreme Court
construed § 337(a) as barring suits by private litigants “for noncompliance with the medical
device provisions.” 531 U.S. 341, 349 n. 4. Thus, to the extent, Plaintiff attempts to raise some
claim, independent of her PLA claim for fraud on the FDA as a result of reporting violations,
such claim would be impliedly preempted. 10 See Mendez, 28 F. Supp. 3d at 291 (“If not
expressly preempted, the state law claim may be still be barred under implied preemption. The
holding in Buckman, however, is often limited to “fraud-on-the-agency” claims and not extended
10
Certain paragraphs of the Amended Complaint, could lend support to the idea that Plaintiff
sought to raise a fraud-on-the-agency claim. See, e.g., Compl. ¶ 46 (“While Boston Scientific
was aware that the Defibrillator might be subject to certain random and infrequent failure,
Boston Scientific was also aware of specific, potentially fatal, and nonrandom failure that would
occur in the device, but failed to disclose any of the subject risks and problems of the device to
the FDA and/or to the public who relies on said device, and failed to take remedial steps to
correct them.”); ¶ 48 (“Defendants, including Boston Scientific and Guidant Corp.’s prior
history, demonstrates that while said defendants had provided some information to the FDA, that
information was incomplete and misleading and did not adequately disclose the device defects.
Boston Scientific’s flawed disclosures did not comply with FDA regulations and violated the
conditions of approval for the devices.”). In the actual Counts of the Amended Complaint,
however, Plaintiff raises only New Jersey PLA and common law theories of liability,
undermining the existence of any fraud-on-the-agency claim.
20
to claims based on state law tort principles.”); Cornett, 211 N.J. at 389 (“Moreover, to the extent
plaintiffs' failure to warn claim is based solely on a contention that defendants obtained FDA
approval for the device only after submitting fraudulent representations to or withholding
material information from the FDA, this claim falls squarely within the Buckman implied
preemption rule. We affirm its dismissal. So, too, plaintiffs' failure to warn claim is preempted
and dismissed to the extent that it can be established solely by evidence of fraud on the
agency.”).
Perhaps acknowledging the great weight of precedent aligned against her position,
Plaintiff, in her Opposition, focuses on a narrow-category of strict products liability failure to
warn cases in which the federal and state courts of this Circuit have found parallel claims not
preempted by the MDA. Citing mostly Mendez, 28 F. Supp. 3d at 291, and cases cited therein,
but relying, without citation, primarily on the reasoning in Cornett v. Johnson & Johnson, 414
N.J. Super. 365, 400 (App. Div. 2010), aff'd as modified, 211 N.J. 362 (2012), 11 Plaintiff argues
that “[w]hen a claim alleging failure to warn related to adverse reactions caused by a medical
device is combined with allegations of nondisclosure, it becomes a claim within a traditional area
of state regulation that would have existed even in the absence of federal requirements. To that
extent, the claims satisfy and warrant avoidance of preemption.” Opp. at 15.
It is concerning that Plaintiff so heavily paraphrased the holding of the Appellate
Division in Cornett without any citation to that case, because a proper citation would have
alerted both the Court and Plaintiff immediately to the fact that the New Jersey Supreme Court
accepted certiorari in Cornett and modified the holding. Specifically, the Appellate Division’s
11
Although the interpretations of federal law by the New Jersey courts are not binding on the
courts of this District, this Court finds the reasoning in the Cornett line of cases persuasive and
discusses them as persuasive authority, infra.
21
holding, although somewhat opaque, at least facially appears to support Plaintiff’s position that a
failure to warn claim coupled with allegations of intentional non-disclosure of information
required by the FDA regulations states a claim not expressly or impliedly preempted by the
MDA. The Appellate Division held that:
The FDCA's rebuttable presumption thus barred a claim about the failure to give such
warnings for approved uses in the absence of deliberate nondisclosure, and such a claim
was also expressly preempted under Riegel as an additional state requirement on
defendants. Hughes, supra, 669 F.Supp.2d at 710; Riley, supra, 625 F.Supp.2d at 780–81.
However, when the claim about the failure to warn for approved uses was combined with
allegations of nondisclosure, it became a claim within a traditional area of state
regulation that would have existed even in the absence of federal requirements. To that
extent, it satisfied Buckman 's test for avoiding implied preemption as a claim that
amounted to no more than “fraud on the FDA.” Hughes, supra, 669 F.Supp.2d at 709–10.
While it is true that the “misconduct” underlying plaintiffs' claims also constitutes a
violation of federal regulations—indeed that is precisely why the claims are parallel—the
suit was brought to vindicate plaintiffs' rights, not the FDA's. Plaintiffs adequately pled
that claim.
Cornett, 414 N.J. Super. at 399–400 (emphasis added).
The New Jersey Supreme Court, however, eloquently disentangled the Appellate
Division’s holding. Under New Jersey law, “Defendants who comply with FDA requirements are
granted a rebuttable presumption that the labeling is adequate.” Cornett, 211 N.J. at 388. “To
overcome this presumption, a plaintiff asserting a failure to warn claim based on an inadequate
label or instructions has stricter pleading requirements. A plaintiff must plead specific facts
alleging deliberate concealment or nondisclosure of after-acquired knowledge of harmful
effects.” Ibid. (quotation omitted). The New Jersey Supreme Court observed that “[t]his pleading
specificity also serves to permit a determination whether a failure to warn claim is preempted by
the MDA or is a permissible parallel state claim.” Ibid. The court explained that the Master
Complaint in Cornett included both state law claims for failure to warn the public about the
dangers of unapproved uses of the device, in that case a stent, which were not part of the PMA
22
process (also called off-label uses) and state law claims for failure to warn of the dangers of
FDA-approved uses. Ibid. The New Jersey Supreme Court found that the plaintiff’s claims for
failure to warn of the dangers of the approved uses of the device were clearly, expressly
preempted under Riegel. The Cornett Court held
The failure to warn claim alleged by plaintiffs includes approved and off-label uses of the
device. As to the approved uses of the . . . stent, the failure to warn claim . . . is nothing
more than a challenge to the adequacy of the information required by the FDA during the
PMA process and label approved by the agency. This failure to warn claim falls within
the PLA rebuttable presumption and the Riegel express preemption rule.
Id. at 389. Those claims were distinguished from the plaintiffs’ unapproved or off-label use
claims for failure to warn which were not preempted. For the unapproved claims, the New Jersey
Supreme Court found that the plaintiffs had adequately alleged “specific conditions of the PMA
of the device and regulations prohibiting promotion of adulterated and misbranded product and
off-label marketing” that had been violated. Id. at 388. Such failure to warn claims concerned
adverse information gathered about the device and parts of the Defendants’ labeling and
instructions that “were not part of the PMA process,” and as such fell “within a traditional area
of state concern and regulation because fraud on the FDA is not an element of the claim and it
can be proved by evidence other than by evidence of fraud on the FDA.” Id. at 390. Accordingly,
the failure to warn claims based on unapproved or off-label uses of the device were not
necessarily preempted. 12
The other case, from this District, upon which Plaintiff principally relies, Mendez, came
to the same conclusion that failure to warn causes of action in off-label promotion cases are not
12
Having so held, the New Jersey Supreme Court observed that even the off-label claims might
be preempted by the safe harbor provisions of the MDA. Cornett, 211 N.J. at 391 (“nevertheless,
if defendants complied with the safe harbor requirements of the MDA in disseminating
information about the off-label uses of the device, plaintiffs’ claim is preempted.”). As Plaintiff
in this case does not raise any off-label claims, no safe harbor analysis by this Court is necessary.
23
necessarily preempted by the MDA. Mendez, 28 F. Supp. 3d at 293 (“plaintiff's claims based on
a theory of off-label promotion of the . . . device is not different from or in addition to federal
requirements, and therefore are not preempted by federal law. . . . Likewise, since plaintiff's
theory is not based on Medtronic defendants committing fraud against the FDA, their off-label
promotion claims are not barred by implied preemption under Buckman.”); id. at 300 (“plaintiff's
failure to warn claim is not preempted” because “[p]laintiff’s theory is based on the idea that
[defendant’s] actions modified the warning label so that the prohibition against using the
[device] with any other component . . . was nullified. Plaintiff is not arguing that [defendant]
should have given warnings that were different from or in addition to the warning provided.”). It
is worthy of note that even though the off-label failure to warn claim in Mendez was not
preempted by the MDA, the district court still dismissed it for failure to clearly state the violated
“federal regulation that parallels [plaintiff’s] state law claim.” Ibid. The off-label claim in
Cornett survived because the plaintiffs had alleged the specific regulations violated.
Here, Plaintiff’s failure to warn claim is not based on unapproved or off-label uses of the
Defibrillator. Instead, Plaintiff clearly asserts that the Defibrillator was defectively designed,
manufactured, and labeled, and the inadequacy of the FDA-approved warnings, which failed to
provide notice of that defectiveness, resulted in injury to the Plaintiff. Such a claim is expressly
preempted by the MDA and does not fall within the category of parallel claims identified in
Cornett and Mendez. Plaintiff simply cannot shoehorn her claim into the mold of a parallel claim
under Riegel, because Plaintiff has failed to allege a violation of FDA regulations and has failed
to allege an independent state law cause of action parallel and not in addition to the federal
requirements that provides a remedy for such violation. Accordingly, Plaintiff’s Count I claim
for strict product liability for failure to warn, along with the common law claims in Counts II-V
24
subsumed within Count I by action of the PLA as explained above, is expressly preempted by the
MDA and is dismissed.
C. Discovery Not Warranted
Lastly, Plaintiff, in her Opposition, requests discovery into Defendants’ compliance with preand-post-PMA approval requirements in order to identify specific violations of FDA regulations
in a future amended complaint. Plaintiff cites no binding authority suggesting that such discovery
would be appropriate. On the contrary, in an unreported opinion, the Third Circuit noted that
“many PMA preemption motions are decided without any discovery.” Smith v. Depuy
Orthopaedics Inc., 552 F. App'x 192, 196 (3d Cir. 2014). The Smith Court went on to hold that
[b]ecause the device components at issue in the products liability action “were approved by the
FDA through the PMA process . . . no discovery was necessary to determine that these
components were also subject to PMA preemption.” Ibid. The courts of this district have thus
consistently found that PMA preemption motions may be resolved before the grant of discovery.
Clements, 111 F. Supp. 3d at 599 (“From the face of the complaint, it is apparent that these
counts, at least as currently alleged, are preempted by the MDA and thus should be dismissed as
a matter of law.”); Becker v. Smith & Nephew, Inc., No. CIV. 15-2538 WHW CLW, 2015 WL
4647982, at *3 (D.N.J. Aug. 5, 2015) (“Plaintiffs contend that they should be permitted to allege
unspecified deviations from FDA requirements at the pleading stage, and fill in the blanks
through discovery. . . . But a plaintiff must successfully plead a claim before obtaining discovery,
not the other way around. Such a premature request for discovery conflicts with Rules 8 and
11(b) of the Federal Rules of Civil Procedure.”); Desai, 2013 WL 163298, at *7 (same).
25
CONCLUSION
For the foregoing reasons, Defendants’ motion to dismiss is granted, and the Amended
Complaint is dismissed.
Dated: _____2/27/2017____
/s/ Freda L. Wolfson
.
The Honorable Freda L. Wolfson
United States District Judge