Almer v. Peanut Corporation of America Doc. 621 Att. 9 EXHIBIT 9 Dockets.Justia.com URGENT MEDICAL DEVICE SAFETY INFORMATION & CORRECTIVE ACTION May 12, 2006 Subject: VITALITY® HE implantable cardioverter defibrillators (ICDs) and CONTAK RENEWAL® 3 and 4 cardiac resynchronization therapy defibrillators (CRT-Ds) implanted beneath the pectoral muscle in an uncommon orientation. Dear Doctor, This letter provides important safety information regarding VITALITY HE ICDs, and CONTAK RENEWAL 3 and 4 CRT-Ds (model numbers listed in Table 1 below). Guidant has received two (2) reports of device malfunction associated with subpectoral implantation in an uncommon orientation (serial number facing ribs). While implant orientation is not reported to Guidant, our records indicate that your health care facility may have implanted or is currently monitoring devices that may be at risk. This advisory is limited to those implanted subpectorally with the serial number facing the ribs (see Figure 2 below). You are encouraged to review the implant positioning for each patient to determine if any of your patients are affected. The vast majority of devices are implanted subcutaneously and are not subject to this failure mechanism. The United States Food and Drug Administration (FDA) may classify this communication as a recall. Description of Issue Two reports of malfunction were associated with subpectoral implantation of devices in certain families of ICDs and CRT-Ds (see Table 1). Accelerated life testing has shown that repetitive mechanical stress applied to a specific area of the titanium case can induce component damage and device malfunction. This can occur only if the device is implanted subpectorally with the serial number facing the ribs. As shown in Figure 2, an anterior/posterior (AP) radiograph can be used to determine device orientation. Anterior view Figure 1. Optimal orientation for subpectoral implants. Leads exit in a counter clockwise direction. Serial number is visible at implant. Figure 2. Avoid this orientation for subpectoral implants. Leads exit in a clockwise direction. Serial number is facing the ribs. Clinical Implications This failure mechanism can result in one or more of the following device behaviors: · Loss of shock therapy · Loss of pacing therapy (intermittent or permanent) · Loss of telemetry communications · Beeping (16 tones every six hours) and a programmer warning screen upon interrogation To date, no patient deaths related to this issue have been reported. One patient required external pacing and immediate device replacement due to lack of pacing therapy. Guidant Corporation 4100 Hamline Avenue North St. Paul, MN 55112-5798 Tel 651.582.4000 Fax 651.582.4166 www.guidant.com Devices Impacted The following model numbers are affected by this communication. Only device families listed in Table 1 that have been implanted subpectorally with the serial number facing the ribs (Figure 2) are subject to this issue. Table 1. Model numbers potentially affected* Device Family Models Device Family Models CONTAK RENEWAL 3 CONTAK RENEWAL 3 HE CONTAK RENEWAL 3 AVT CONTAK RENEWAL 3 AVT HE VITALITY HE H170/H173/H175 H177/H179 M150/M155 M157/M159 T180 CONTAK RENEWAL 4 CONTAK RENEWAL 4 HE CONTAK RENEWAL 4 AVT CONTAK RENEWAL 4 AVT HE H190/H195 H197/H199 M170/M175 M177/M179 *Not all devices are approved in all geographies Due to component location, damage associated with this failure mode will not occur in a subcutaneous position. Similarly, the failure will not occur in devices implanted in a subpectoral position with the serial number facing up (Figure 1). Rate of Occurrence The implant orientation of devices is not reported to Guidant. For this reason, no rate of occurrence or failure rate projection can be provided. However, based on available information, Guidant estimates that the number of devices implanted in a susceptible orientation is likely less than 1% of the total population. The following factors may also impact the risk of failure: · · · · Exact location of the patient's ribs relative to the pulse generator case Body size or muscle mass of the patient (risk increases for larger patients) Activity level or occupation of the patient (risk increases for more active patients) Duration of implant (the risk of component damage increases as devices age, due to repetitive rib stress) Recommendations For patients implanted with a model listed in Table 1, review records to determine if the device is subpectoral. 1. For patients with a subpectoral implant, use an AP radiograph to determine device orientation. If the device is in a susceptible orientation, the patient should be advised of the potential for device failure, and patient follow-up at least once per quarter is recommended in accordance with device labeling. Consider device replacement for muscular or active patients or for patients who regularly need device therapy. If the patient's device was implanted subpectorally and leads exit the pulse generator in a counter clockwise direction as shown in Figure 1, no change to your current patient management is necessary. A list of devices specific to your clinic that may be at risk if implanted in a susceptible subpectoral orientation is available from your local Guidant sales representative. A device model and serial number search tool, available at www.guidant.com in the Product Safety Information section, will be updated as soon as possible. 2. For future implants, when using subpectoral implantation of a device from any product family listed in Table 1, orient the device with the serial number facing away from the ribs, as shown in Figure 1. Warranty Supplement Program Guidant's warranty supplement program, subject to certain conditions, provides a no cost replacement device and up to $2500 in unreimbursed medical expenses. Further Information Guidant will provide additional information as it becomes available. We recognize the impact of this communication on both you and your patients, and want to reassure you that patient safety remains our primary concern. As always, if you have any questions regarding this communication, please contact your local Guidant representative or Guidant Technical Services at 1.800.CARDIAC (227-3422). Sincerely, Renold J. Russie Director, Product Performance Reporting Guidant Cardiac Rhythm Management

Disclaimer: Justia Dockets & Filings provides public litigation records from the federal appellate and district courts. These filings and docket sheets should not be considered findings of fact or liability, nor do they necessarily reflect the view of Justia.


Why Is My Information Online?